Heath-consultants GT Series Manuale Utente Pagina 81

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 170
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 80
7-5
CO (CARBON MONOXIDE) MODE
7.4.3 Air-Free CO
Air-free CO is the CO reading modified by the O
2
reading,
therefore this measurement will only be available when
an O
2
sensor is fitted. The purpose is to determine whether
or not the emissions from an un-ventilated appliance,
principally an oven or a stove / cooker, are safe.
This is the CO reading modified by the O
2
reading
(only applicable if O
2
sensor is fitted).
CO Air-free = (20.9 x CO) / ( 20.9 O
2
).
The typical display for this mode is shown in Fig. 7.5:
Fig. 7.5 Air-Free CO
A sample is considered invalid until the Oxygen (O
2
)
concentration is below 19% Vol.
Changes to Fig 7.6 when valid sample is taken
Fig. 7.6 Valid Sample
Vedere la pagina 80
1 2 ... 76 77 78 79 80 81 82 83 84 85 86 ... 169 170

Commenti su questo manuale

Nessun commento